Laboratory Medicine
Test catalogue
Osmolality, Urine Random | |
---|---|
Test NameOsmolality, Urine Random |
|
Specimen Type / RequirementsContainer Type = Sterile Container Minimum Testing Volume = 5ml Urine |
|
Specimen HandlingCollection Instructions: Random urine collected in red top vacutainer or sterile urine container, Ensure both sample and requisition are labelled properly. Write your initials, date and time of collection on the collection label. Stability:Stable for 1 week at 2-8°C.. Transportation:Transport at room temperature ASAP. Laboratory Staff Instructions: No aliquoting is necessary for urine Osmolality. Samples are analyzed by the gas bench. Sample Rejection Criteria: Unlabeled, leaking and/or Non sufficient quantity (Not Sufficient Quantity) will be rejected. |
|
Required Documentation |
|
Turnaround Time (TAT)45 Mins for Trauma / Stat - 2 hours for routine |
Test UtilizationRoutine |
Reference ValueUrine 50 – 1200 mOsm/kg |
|
Test CodeUOSM |
MethodologyFreezing-point depression osmometry |
Testing LocationSMH-Biochemistry |
Other InformationLast Updated: February 10, 2020
|